top of page
< Back
Daviteq GHC Flammable Gas Sensor
LoRaWAN
Sigfox
Sub-GHz
NB-IoT

Daviteq GHC Flammable Gas Sensor

GHC

Sản phẩm thay thế mới

SKU được thay thế

Satellite IoT

Phòng trưng bày

1. Giới thiệu

1.1 Overview

Daviteq GHC Gas sensor module is intended for the automatic continuous measurement of hydrocarbons or carbon dioxide concentration in the atmosphere. The sensor operating principle is based on NDIR technology, i.e. on selective absorption of LED-produced infrared radiation by gas molecules. The differential dual wavelength method allows the elimination of water vapor, optical elements contamination, and other non-selective hindrances influence.

It has ultra-low power consumption to allow it to be integrated with Wireless Devices such as Sub-GHz transmitter, Sigfox transmitter, LoRaWAN transmitter, RS485 output transmitters, etc.

It can be calibrated with target gas like CH₄, C₃H₈, CH₄, C₂H₆...

Typical Applications: Flammable gas detection.

1.2 Specification



Detailed reading stability




1.3 Cross-Sensitivity Data


What is cross-sensitivity?

The gas detection sensor is usually affected by other gas. It meant the sensor not only measure the target gas but also the other gases. If there is a concentration of additional gas, it would also cause a change in sensor output with a factor listed in the below table.



2. Nguyên tắc hoạt động

When infrared radiation interacts with gas molecules, infrared light is absorbed by the gas molecules at a particular wavelength, causing vibration of the gas molecules. NDIR (Non-Dispersive Infrared) gas sensors detect a decrease in transmitted infrared light which is in proportion to the gas concentration. This transmittance, the ratio of transmitted radiation energy to incident energy, is dependent on the target gas concentration.


NDIR gas sensor consist of an infrared source, detector, optical filter, gas cell, and electronics for signal processing. A single light source, dual wavelength type gas sensor has two detectors and two optical filters of different wavelengths which are placed in front of each detector. The infrared light that is absorbed by a target gas passes through the active filter with a particular bandwidth for the detection of the target gas. The infrared light that does not interact with the target gas passes through the reference filter. The difference between transmitted light intensities in these two bandwidths is converted into gas concentration. The dual wavelength sensor ensures stable measurements for a long period of operation as the aging effects of the light source or the gas cell are automatically compensated by output signals at the reference wavelength.

 

Mid-infrared radiation through sample gas causes a resonance of gas molecules at their natural frequency with the infrared light in the spectrum region where the energy level of infrared is equivalent to the natural frequency of gas molecules, resulting in the absorption of infrared by gas molecules in the form of molecular vibration.

The relationship between infrared transmittance and gas concentration is expressed by the Lambert-Beer law:



Where T is transmittance, I is the intensity of light passed through sample gas and an optical filter, Io is the initial light intensity emitted from the source, ε is the molar attenuation coefficient, c is gas concentration, and d is the light path length.

Because ε of the target gas and the light pass length d are fixed with an NDIR sensor, gas concentration can be measured by measuring the transmittance within the spectrum region of the absorbed energy (wavelength) by the target gas.


The initial light intensity emitted from the light source Io is preset by calibration using zero gas which does not absorb infrared light. The initial value of the molar extinction coefficient ε is set by calibration using calibration gas of known concentration.

3. Hiệu chuẩn

The Daviteq GHC Gas Sensor must be connected to a reading device, it usually is a wireless transmitter like Sub-GHz, Sigfox, or LoRaWAN.

Why do we need to calibrate the gas sensor? There are some reasons:

- The output value of a sensor is different from the other sensor. It is not the same value for all sensors after manufacturing.

- The output value of a sensor will be changed over time.

Therefore, users need to calibrate the sensor before use or in a pre-defined interval time (30 months for example).

How to calibrate the GHC Flammable Gas sensor?

NOTE: THE CALIBRATION CAN ONLY BE DONE IN THE SAFE ZONE!!!


Please follow steps for Instruction to attach the calibration cap onto the sensor module to get Zero or Span values:


Step 1. Remove the Filter and prepare the calibration cap



Step 2. Attach the calibration cap to the sensor head


Step 3. Installed the Regulator to the Gas cylinder


Step 4. Attach the tube to the regulator

Please select the flow regulator with a flow rate of 2.5 LPM or 5.0 LPM.


With the 2-point calibration method, the user can define the A and B factors. Please find below the steps of calibration.


Step 1. Get the Zero value.

- Power ON the device;

- Place the device in a clean-air environment (the target value is nearly zero) at a temperature from 20∼30℃, in at least 60 minutes.

- After 60 minutes, force the device to send data, read and record the Raw_value, so now you got the Zero_value = Raw_GHC value.

Recommendation: Record many Raw_GHC values ​​at least 10 minutes apart (10 values). Zero value is the average of the recorded Raw values.


Note: The Raw_GHC values can be positive or negative; Its value is usually 7 (% LEL)


Step 2. Get the Span value


Note: Keep the sensor Power ON all the time


- Use the standard gas cylinder with a known concentration (for example Eythylene Air  1.35% is equivalent to 50% LEL ) to supply the gas to the sensor;

- Use the calibration cap as above pictures to attach to the sensor and connect the tubing to the gas cylinder;

- Open the valve on the Cylinder slowly and make sure the gas has reached the sensor. The flow regulator should be 2.5 LPM or 5.0 LPM.


Notes: The tube length is short as possible to reduce the gas loss.


- Press a timer to start counting the time;

- After 2 minutes, force the device to send data once every minute, and stop forcing at 5 minutes;

- The highest Raw_value is the Span value.


Note: Just get one value for Span.


- After that, immediately turn OFF the valve to save the gas;

- Remove the calibration cap from the sensor;

- Place the sensor in clean air again.


Note: Always keep the sensor Power ON all the time.


Step 3. Calculate the new A and B

The calculation of new A, B value based on basic linear formula:

y = A * x + B

Where:

A, B is calibration coefficients

x is the sensor process value (example gas level in ppm) read on reading device such as on application server/network server, on offline tool. The process value is the RAW_VALUE in the payload

y is the correct value. y is the value of standard gas/standard condition


Which condition of Zero value: y₀ = A * x₀ + B

Which condition of Span value: yₛ = A * xₛ+ B


From the two formulas, the calculation of A, B as below

A = (y₀ - yₛ) / (x₀ - xₛ)

B = (yₛ * x₀ - y₀ * xₛ) / (x₀ - xₛ)


Example of A, B calculation for LoraWAN Ammonia Gas sensor (item code WSLRW-G4-NH3-100-01):

* With condition of clean-air environment at a temperature from 20∼30℃, there is no ammonia gas (y₀ = 0); while the NH₃ level on reading device (RAW_VALUE in the payload) is -0.25 (x₀ = -0.25)

* When the sensor is connected to standard gas cylinder having ammonia level of 25 ppm (yₛ = 25); while the NH₃ level on reading device (RAW_VALUE in the payload) is 18.66 (xₛ = 18.66)


The calculation of A, B for the Ammonia gas sensor:          

A = (0 - 25) / (-0.25 - 18.66) = 1.32205

B = (yₛ * x₀ - y₀ * xₛ) / (x₀ - xₛ) = (25 * (-0.25) - 0 * 18.66) / (-0.25 - 18.66) = 0.33051


The factory default A = 1 and default B = 0


The RAW_VALUE in the payload is used for calibration


Step 4. Configure the new A and B into the device

- User can use the off-line tool or downlink to write the values of A and B;

- Writing the new A and B successfully meant you had done the calibration process. Congratulation!

4. Ghi chú ứng dụng

5. Installation Notes

Notes: If a sensor has been kept in transport containers at temperatures below zero centigrade, leave it at 1035 for not less than one hour.* if the Sensor is intended to install outdoors, please use a rain guard to protect the sensor from rain and direct sunlight. Please contact us to buy this accessory.

  • Place the sensor in the area to monitor the target gas concentration. Please always check the gas molecular weight v.s the air.

  • Do not use a damaged sensor. It must be repaired only by personnel authorized by the manufacturer.

  • Keep the sensor out of contact with aggressive substances, e.g., acidic environments, which can react with metals, and solvents, which may affect polymeric materials.

  • Diffusion holes of the sensor should be protected against the ingress of sprayed liquid or waterdrops, buy using the Splash guard or Rain guard.

  • The sensor is not intended to measure the target gas concentration contained in fluids.

  • Correct measurement is provided when the ambient temperature changes not faster than 0.6/ min.

  • Inspection and maintenance should be carried out by suitably trained personnel.

  • Persons, who have studied this guide, must be briefed on safety precautions when operating electrical equipment intended for use in explosive areas in due course.

  • When dealing with a cylinder containing a gas mixture under pressure, it is necessary to follow safety regulations.

  • There is no risk of pollution or negative impact on human health. The sensor does not contain any harmful substances that may be released during its normal operation.

6. Troubleshooting



7. Maintenance



Sensor replacement instructions:


Please remove the batteries before doing the following steps.


Step 1. Unscrew the gas housing


Step 2. Remove the filter housing


Step 3. Unplug gently the sensor module


Step 4. Plug gently and firmly the new sensor module into the PCB


Step 5. Insert the batteries and start calibration as per Section 3


Step 6. Place the filter cover back and screwing to lock


8. Cấu hình mặc định

This GHC gas sensor module has the default configuration, however, those parameters can be changed. The user can change the configuration on the wireless transmitter so that the complete sensor (transducer + wireless) delivers the proper output value. Below are some configuration parameters that store in the flash memory of the wireless transmitter.



bottom of page